Measurements of Stimulator (IR Source)

Darren Dowell -- (626)395-6675 (office), -2600 (lab), 796-8806 (FAX)


Last modified Wednesday, 11-July-2001 9:48 PDT
cdd@submm.caltech.edu

Radiation and Electrical Properties -- July 2001

Devices

The stimulator under test was provided by M. Jhabvala (NASA-GSFC) and is of the type used for SIRTF. We tested die 060-055.

Conclusions

When heated to 100-300 K, the stimulator produces plenty of signal in a broad infrared band for detection by a bolometer. We observed on the order of 1-10 pW detected by a 1 mm x 1 mm bolometer located 40 mm away. The time constant (1/e-folding time) of the stimulator is < 1 msec and is therefore useful for measuring bolometer radiation time constants > 1 msec.

It is unclear (untested) whether the stimulator produces enough signal in a narrow far-IR band to be useful for detector absorptivity measurements.

Experimental Set Up

The measurements were performed in the Barney cryostat. During the tests with bolometers exposed to unfiltered radiation from the interior of the cryostat, we observed a radiation leak of approx. 60 pW per detector with the stimulators off. The radiation leak was stable over two cooldowns and did not interfere with the stimulator measurements reported here.

back view of stimulator front view of stimulator close-up view of stimulator

The bolometer array used for the measurements was 5327 LH5, a "G0 assortment pack" from the Spring 2000 calibration run. The array was recently coated with a SHARC II-like bismuth absorbing film. The bolometers were approx. 40 mm from the stimulator.

Electrical Properties

The stimulator was successfully driven with both constant current and constant voltage excitation. The resistance of the device has a strong temperature dependence as shown below:
R vs. T
Figure. Resistance of device 060-055 vs. temperature. The excitation was 1 mV at high temperatures and 100 microV at low temperatures and was observed to produce negligible heating in the device. The green curve shows measurements by D. Robinson (NASA-GSFC) for a separate device. Heating by the 10 microA current excitation is evident at low temperatures.

The resistance of the device is used to estimate the temperature under load.

The electrical time constant is measured by exciting with a constant current and observing the settling of the voltage (or vice versa). For heating from 4 K to approx. 300 K, the electrical time constant was < 0.4 msec.

Radiation Produced

The stimulator and bolometer array were cooled down twice -- once with no filter, and once with a silicon wafer (396 microns thick) in front of the stimulator. Silicon should block all optical light and some bands in the near and mid infrared. Measurement details are available. With an excitation of 1.4 V, which heated the device to approx. 300 K, signals of ~18 pW and ~5 pW were observed in the unfiltered and filtered cases. With an excitation of 0.58 V, which heated the device to approx. 80 K, signals of ~0.42 pW and ~0.30 pW were observed in the unfiltered and filtered cases. When the stimulator was warmer, a greater fraction of the radiation was blocked by the silicon, indicating the expected shift in the spectrum toward shorter wavelengths.

Radiation Time Constants

Both radiation and electrical time constants were performed. For the type 3A bolometers with large bias voltage, radiation time constants of approx. 1.0 msec were observed, verifying the speed of the stimulator.

tau vs. bias

radiation vs. electrical

Examples of measurements:
radiation/slow slow radiation time constant radiation/fast fast radiation time constant
electrical/slow slow electrical time constant electrical/fast fast electrical time constant


Go to SHARC II home page...