Stimulator 060-055 #1 Bolometer array 5327 LH5, Bi absorber, 308 mK, bias = 392 mV through 30 Mohms Measured July 2001 at Caltech, two separate cooldowns with and without silicon filter (396 microns thick) Stimulator Signal(mV,pW) Signal(mV,pW) Bolometer Excitation (mV) No filter Si filter ----------- --------------- ------------- ------------- 16 type 3A 1740 (0.86 mW) 0.899 18.2 16 type 3A 1620 (0.67 mW) 0.484 9.82 16 type 3A 1600 0.505 10.2 16 type 3A 1400 0.937 19.1 0.258 5.23 15 type 3A 1400 1.085 22.2 0.266 5.41 14 type 3A 1400 0.858 17.7 0.273 5.62 4 type 2B1 1400 0.688 17.8 0.187 4.78 3 type 2B1 1400 0.608 15.7 0.179 4.59 2 type 2B1 1400 0.562 14.6 0.169 4.34 1 membrane 1400 0.923 13.9 0.260 3.83 16 type 3A 1000 0.0688 1.40 16 type 3A 800 0.0347 0.70 16 type 3A 646 (0.06 mW) 0.0164 0.33 16 type 3A 580 0.0199 0.41 0.0137 0.28 15 type 3A 580 0.0217 0.44 0.0154 0.31 14 type 3A 580 0.0198 0.41 0.0120 0.25 4 type 2B1 580 0.0166 0.43 0.0117 0.30 3 type 2B1 580 0.0144 0.37 0.0104 0.27 2 type 2B1 580 0.0152 0.39 0.0084 0.22 1 membrane 580 0.0200 0.30 0.0134 0.20 16 type 3A 400 0.0060 0.12