Pop-Up Detector Test Report

26 April 2000 11AM
C. Darren Dowell
Caltech, Mail Code 320-47
Pasadena, CA 91125
(626)395-6610 (office)
(626)395-2600 (lab)
(626)796-8806 (FAX)
cdd@submm.caltech.edu

Abstract

Measurements of semiconducting pop-up bolometers were performed at Caltech in April 2000. We report the resistance vs. temperature and the thermal characteristics of the measured samples. The desired SHARCII and HAWC parameters were bracketed with the measured samples, and we make recommendations for the bolometer designs for the two instruments.

Acknowledgments

Walter Collins (Caltech), Mino Freund (Goddard), Matt Gardner (Caltech), and Jeff Groseth (Caltech) helped considerably with preparing the test components and in performing the measurements. Attila Kovacs (Caltech) wrote the software to solve for bolometer parameters from measured data. Christine Allen (Goddard) provided the bolometer samples and advice for handling them. Harvey Moseley (Goddard) provided guidance through the testing period.

Outline

  1. Measurement apparatus
  2. Bolometer design
  3. Thermistor results
  4. Results for 5327 G0 test array
  5. Results for 5347 G0 test array
    1. Fitted parameters
    2. White noise
    3. Time constants
  6. Results for 5347 thermistor test array
  7. General conclusions
  8. Recommendation for SHARCII bolometer design
  9. Recommendation for HAWC bolometer design

Measurement Apparatus

Implant and bolometer samples were tested in the SHARCII cryostat, which contains a 3He refrigerator enclosed in a L4He-cooled radiation shield. For most measurements, the main L4He reservoir was pumped down to 1.5 K so that the device substrate reached temperatures as low as 0.325 K.

implant test devices bolometer test devices package with cover on
Figure 1. Interface substrate. The 3-element implant screening devices were GE varnished to the substrate, and the 32-element bolometer arrays were clipped to the substrate. Gold wedge bonds provided electrical connection between the devices, load resistors, and substrate. Surface-mount connectors were initially used to interface with the cryostat wiring. However, they were replaced with directly soldered wires for later measurements. The ceramic substrate was partially enclosed in an INVAR box. A sheet of gold leaf and the force of ~6 screws were used to thermally connect the INVAR and ceramic.

assembly, shield off assembly, shield on
Figure 2. Device suspension and shielding. The ceramic/ INVAR package was suspended from the L4He coldplate with 3-4 fiberglass tubes with A/L = 0.014-0.018 cm. The package was heat sunk to the fridge through an annealed OFHC copper strap with A/L = 0.014 cm. A calibrated LakeShore GRT was attached to the ceramic substrate with a screw and gold leaf, approximately 5 cm from the detectors. The package was shielded by a secondary L4He enclosure with blackened walls. With this configuration, the coldest observed GRT measurement for the detector package was 0.325 K, during which the L4He bath was at 1.5 K and the fridge was at 0.291 K.

device on coldhead
Figure 3. Additional implant screening hardware. In order to confirm the implant resistances, the 3-element devices in the Goddard- selected ceramic package were attached to a copper block and installed directly on the SHARC II fridge coldhead or in a separate L4He cryostat (the purple 'Barney' Dewar).

All wires going into the Dewar pass through one or two RFI-filtered connectors, which contain 4 nF capacitors from each pin to the Dewar wall.

***Thermal properties of detector package, gold bonds, and silicon frame. The silicon frame has a thickness of 300 microns. In the various directions of heat flow, the relevant A/L is from approximately 0.0015 cm to 26 cm. For a 0.3 K thermal conductivity of *** The wirebonds were 0.001 inch diameter 99.99% gold. With 64 each 3 mm long wire bonds (total A/L=0.001 cm), and assuming a 0.3 K thermal conductivity of 0.1 W/cm/K (Touloukian et al. 1970), the effective G of the wirebonds was 100 microW/K. We have ignored unknown boundary thermal resistance, which will reduce the effective G. The ceramic substrate... The heat strap also limits the cooling of the detector wafer. During the cooldowns reported here, the fridge/heat strap combination was observed to have an effective G of 80 microW/K at the coldest temperatures.

For a 0.5 V bias across the 30 Mohm load resistors (and detectors), the dissipated power can be at most 0.5 microW.


Designs of Test Bolometers

Three types of bolometer arrays were manufactured for the detector measurements:

G0 recipes
Table 1. Recipes for bolometers on G0 test arrays.

thermistor recipes
Table 2. Recipes for bolometers on thermistor test arrays.

thermistor recipes
Table 3. Implant doses. Higher net density means greater implant dose, which means lower cold resistance. Wafers 5347 and 5327 are of interest for SHARCII and HAWC.


Thermistor Results

The first part of the detector measurement program was to screen five wafers for desired resistance at cold temperatures. Three-element membrane devices were measured in three configurations -- GE varnished to the Caltech ceramic interface board, packaged in the Goddard ceramic carrier on a copper block in the purple L4He dewar, and in the Goddard carrier/copper block on the SHARCII coldhead.

Resistance measurements were made with a Keithley 616 Digital Electrometer borrowed from Goddard that allows current excitations from 1 pA to 1 microA stepped by factors of 10. The reported measurements were made with the highest current setting which does not cause significant self heating. The criterion was to allow no more than a 5% effect on the observed resistance.

SHARCII Dewar, Large Ceramic Board

SHARCII Dewar implant tests
Figure 4. Resistance measurements for 3-element membrane devices on Caltech ceramic board. The SHARCII target curve is T0 = XXX, R0 = YYY (REFERENCE), and the HAWC target is T0 = 30 K, R0 = 700 ohm (Moseley 1999).

SHARCII Dewar implant tests 2
Figure 5. Resistance measurements plotted in units which give linear fits.

A fit to the data with model R = R0 exp[(T/T0)1/2] yields:

Label   ND   R0 (ohms)  T0 (K)
-----  ----  ---------  ------
5250a  0.6     2570      81.8
5250b  0.6     3880      74.6
5327   0.7     1750      48.9
5347   0.75    2090      25.7
Table X. CAPTION... Checked these.

SHARCII Dewar, Small Ceramic Package on Coldhead


Purple Dewar, Small Ceramic Package

Three-element dies were tested using the Goddard-provided ceramic packages attached 'upside down' to a copper block in order to shield the devices from radiation. Resistances were measured with the Keithley 616 electrometer.

purple Dewar implant tests purple Dewar implant tests 2 Figure X. CAPTION... Checked these.

A fit to the data from the purple Dewar experiment only yields:

Label   ND   R0 (ohms)  T0 (K)
-----  ----  ---------  ------
5306   0.65    1670      67.3
5327   0.7     2100      42.7
5347   0.75    2380      23.2
Table X. CAPTION... Checked these.

Purple Dewar, small ceramic package


Results for 5327 G0 Test Array LH5

***Circuit. JFET gain.

The bolometer parameters are derived from a minimization procedure in which the measured I's and V's are compared with a best-fit four-parameter model. The model is:

See Mather (1982) for further details of the model.

Table X.  Four-parameter fits for G0 bolometer
array 5327 ***.  The four parameters are R0, T0, G0, and beta.


Parameter Fits for 5347 G0 Test Array LH1

All bolometers were wired with 30 Mohm nichrome load resistors from MSI. They were measured with two separate cooldowns between which the routing of the 16 JFETs was changed.

IV's for 5347 LH1 bolometer 23
Figure X. Sample IV curve for a single bolometer at multiple temperatures. The filled circles are measured data, and the lines are model fits with 4 free parameters.

G0 test array 5347 LH1
Data at 330, 440, 597, 742, 985, 4243 mK (pixels 1-16)
Data at 329, 448, 604, 744, 974, 4261 mK (pixels 17-32)
Updated 26 Apr 2000, 10:30 PDT

R = R0 exp(sqrt(T0/T))
G = G0 T^beta

                        R*     R
          R0    T0    0.3 K  0.5 K   G0=G(1 K)          G(0.3 K)*  G(0.5 K)
bol/grp  ohms    K    Mohms  Mohms  W/K^(beta+1)  beta     W/K       W/K
-------  ----  -----  -----  -----  ------------  ----  ---------  ---------
01 mem.  1844  26.25   21.3   2.59    23.3 e-9    0.55  12.1  e-9  15.9  e-9
02 TH.1  1290  25.58   13.2            1.07e-9    1.32
03 TH.1  1365  25.51   13.8   1.73     1.20e-9    1.19   0.29 e-9   0.53 e-9
04 TH.1  1401  25.31   13.7            0.94e-9    1.15
05 HG3D  1953  26.94   25.5           24.4 e-9    1.97
06 HG3D  1956  27.16   26.5   3.11    24.1 e-9    1.96   2.28 e-9   6.19 e-9
07 HG3D  2147  27.11   28.9           19.4 e-9    1.77
08 HG3C  2127  26.93   27.7            2.24e-9    1.20
09 HG3C  1844  27.58   26.9   3.10     2.66e-9    1.38   0.51 e-9   1.02 e-9
10 HG3C  1898  27.05   25.2            2.61e-9    1.35
11 HG3B  1934  28.40   32.5           24.2 e-9    2.03
12 HG3B  1908  28.51   32.7   3.63    25.1 e-9    2.06   2.10 e-9   6.02 e-9
13 HG3B  1869  28.76   33.4           27.8 e-9    2.14
14 HG3A  1840  28.67   32.4            4.08e-9    1.44
15 HG3A  1951  28.41   32.9   3.66     3.80e-9    1.42   0.69 e-9   1.42 e-9
16 HG3A  1924  28.37   32.2            3.89e-9    1.41
17 HG2D  1893  28.73   33.7           38.4 e-9    2.63
18 HG2D  1899  28.90   34.8   3.80    38.7 e-9    2.66   1.57 e-9   6.12 e-9
19 HG2D  1981  28.82   35.8           36.7 e-9    2.65
20 HG2C  2025  28.51   34.7            1.17e-9    1.35
21 HG2C  1839  28.97   34.1   3.72     1.31e-9    1.43   0.23 e-9   0.49 e-9
22 HG2C  1920  28.91   35.2            1.26e-9    1.40
23 HG2B  1957  29.47   39.4            1.21e-9    1.37
24 HG2B  2038  29.15   38.9   4.22     1.17e-9    1.34   0.23 e-9   0.46 e-9
25 HG2B  1929  29.84   41.4            1.25e-9    1.40
26 HG2A  1871  30.40   44.0           42.4 e-9    2.76
27 HG2A  2019  29.93   44.0   4.63    37.3 e-9    2.62   1.59 e-9   6.07 e-9
28 HG2A  1952  30.41   46.0           40.0 e-9    2.69
29 HG1A  1836  31.16   49.0           20.0 e-9    2.74
30 HG1A  1788  31.95   54.2   5.30    20.7 e-9    2.80   0.71 e-9   2.97 e-9
31 HG1A  1927  31.34   52.9           18.1 e-9    2.68
32 mem.  1814  32.12   56.5   5.49    72.9 e-9    1.58  10.9  e-9  24.4  e-9

* extrapolation
Table X. Four-parameter fits for G0 bolometer array 5347 LH1.

***Interpretation...

Resistance Gradient

The cold resistance goes from high at bolometer 32 to low at bolometer 1. In order to confirm this situation, a few bolometers were measured individually with an electrometer. The 0.327 K resistance measured in the linear portion of the IV curve is tabluated below:
           Electrometer  Table X Fit
            Resistance   Resistance
Bolometer      Mohm         Mohm
---------  ------------  -----------
    1          15.6         14.4
    9          19           18.0
   12          23           21.7
   15          22           21.8
   24          28           25.7
   32          36.4         36.6
*** Do calculation for power required to warm bolometer 1 up from resistance of bolometer 32.

Noise for 5347 G0 Test Array LH1

***Circuit.

Time Constants for 5347 G0 Test Array LH1

***Circuit.

For the time constant measurements, the bolometers were driven using a bias waveform with a square profile switching between two positive levels 8 mV apart. The output of the JFET was sent to an SR560 preamplifier with a gain of 100 and then to an oscilloscope. DC coupling was used on both instruments; a DC level from a power supply was subtracted to bring the signal on scale. Low-pass filtering was applied with the SR560, but the cutoff frequency (typically 3 kHz) was chosen so that only high frequency noise was eliminated and the shape of the waveform was preserved.

Additional filtering of the bolometer waveform was caused by the bias waveform filtering (cutoff frequency at 6800 Hz) and the RFI connectors (cutoff frequency at 20 kHz, assuming a 1000 ohm JFET output impedance). However, these cutoff frequencies are high enough to be irrelevant.

tau: bol. 24, 40 mV
Figure ***. Typical oscilloscope trace during time constant measurement. Shown is the output of the JFET for bolometer 24 of the 5347 G0 test array, multiplied by a gain of 100. The bias frequency was 8 Hz, and the settling time was 3.6 msec.

The overshoot is caused by the high-frequency impedance of the bolometer being larger than the low-frequency impedance (Mather 1982). Stated another way, the dynamic circuit model of a bolometer contains an effective inductance, which creates a voltage spike at the transition of the bias. Our interpretation is that the overshoot spike is rolled off (i.e., not instantaneous) by parasitic capacitance (time constant Rbolometer x Cparasitic). The settling of the voltage following the bias transition is the true detector time constant (taue in the notation of Mather 1982).

*** Reconsider overshoot tau -- note large t(o.s.) for low G (bol. 24).

For the measured quantities, we looked at the transition caused by the positive change in the bias. We recorded:

In the table below, voltages are referred to JFET output, before the preamplification. The calculated G and T are the estimates from the IV curve fits.
                                                     calc.   observ.
            bias  calc G  calc T  V(o.s.)  t(o.s.)  V(step)  V(step)  t(settle)
bol.  grp.   mV    nw/K      K       mV      msec      mV       mV       msec
----  ----  ----  ------  ------  -------  -------  -------  -------  ---------
  9    3C    25     0.59   0.34     0.12     2.2      2.9      2.7       1.5
  9    3C    49*    0.64   0.36     0.42     1.1      1.4      1.4       1.6
  9    3C    98     0.76   0.40     0.70     0.52     0.34     0.40      1.2
  9    3C   196     0.98   0.48     0.61     0.28     0.02     0.06      0.71

 15    3A    39     0.84   0.34     0.19     1.8      2.2      2.2       1.6
 15    3A    78?*   0.95   0.38     2.8?     0.92     0.90     6.6?      1.1 
 15    3A   157     1.20   0.44     0.69     0.31     0.12     0.18      0.61
 15    3A   314     1.60   0.54     0.47     0.24    -0.02     0.05      0.31

 18    2D    49     2.1    0.34     N.A.     N.A.     2.7      2.6       0.38
 18    2D    98*    2.5    0.36     0.20     0.8      1.5      1.5       0.68
 18    2D   196     3.6    0.41     0.46     0.31     0.45     0.53      0.36
 18    2D   392     5.6    0.48     0.34     0.22     0.12     0.22      0.16

 24    2B    20     0.28   0.34     0.39     2.4      2.6      2.7       4.5
 24    2B    39*    0.31   0.37     1.1      1.3      1.2      1.1       3.6
 24    2B    78     0.38   0.43     1.2      0.6      0.17     0.11      2.3
 24    2B   157     0.50   0.53     0.74     0.4     -0.03    -0.13      1.3

* optimum bias for NEP
? apparent errors in measurement, probably bias amplitude setting

Results for 5347 Thermistor Test Array RH7

Fitted Parameters

Thermistor test array 5347 RH7
Data at 330, 451, 610, 755, 971, 4190 mK
Updated 28 Apr 2000, 09:45 PDT

R = R0 exp(sqrt(T0/T))
G = G0 T^beta

                        R*     R
          R0    T0    0.3 K  0.5 K   G0=G(1 K)          G(0.3 K)*  G(0.5 K)
bol/grp  ohms    K    Mohms  Mohms  W/K^(beta+1)  beta     W/K       W/K
-------  ----  -----  -----  -----  ------------  ----  ---------  ---------
09 TH.1  1622  35.93   91.8            1.53e-9    1.67
10 TH.1  1658  35.50   87.9   7.57     1.52e-9    1.64   0.21 e-9   0.49 e-9
11 TH.1  1622  35.65   88.0            1.54e-9    1.69
12 TH.2  1800  34.29   79.1            1.42e-9    1.49
13 TH.2  1733  34.60   80.0            1.48e-9    1.53
14 TH.2  1726  34.57   79.3   7.05     1.49e-9    1.52   0.24 e-9   0.52 e-9
15 TH.2  1800  34.52   82.0            1.44e-9    1.51
16 TH.2  1802  34.41   80.7            1.44e-9    1.50
17 TH.3  1638  36.36   99.0            1.50e-9    1.57
18 TH.3  1668  36.19   98.2            1.49e-9    1.56
19 TH.3  1727  36.17  101     8.53     1.43e-9    1.56   0.22 e-9   0.48 e-9
20 TH.3  1693  36.09   98.2            1.46e-9    1.56
21 TH.3  1578  36.79  102              1.55e-9    1.62
22 TH.4  1096  33.11   40.0            1.42e-9    1.45
23 TH.4  1106  33.47   42.8   3.95     1.37e-9    1.42   0.25 e-9   0.51 e-9
24 TH.4  1108  33.03   40.0            1.37e-9    1.42

* extrapolation
Table X. Four-parameter fits for thermistor test array 5347 RH7.

There is no strong resistance gradient as on the LH1 G0 device, as shown in the following 0.330 K measurements acquired with an electrometer:

           Electrometer  Table X Fit
            Resistance   Resistance
Bolometer      Mohm         Mohm
---------  ------------  -----------
    1          15.6         14.4
    9          19           18.0
   12          23           21.7
   15          22           21.8
   24          28           25.7
   32          36.4         36.6

Low Frequency Noise

Differential AC-biased readout
Figure X. CAPTION

Time Constants


General Conclusions


Recommendation for SHARCII Bolometer Design


Recommendation for HAWC Bolometer Design


References

Mather, J. C. 1982, Applied Optics 21, 1125, "Bolometer noise: nonequilibrium theory"

Moseley, H. 1999, memo dated July 29, "Detector Design for HAWC"


**************************************************************

Thermal Conductance Variety Pack 5327 -- Results -- Apr. 12, 2000

Tests of G0 array LH5 with 5327 (ND=0.7) doping are complete. Bolometers 17-29 and 32 were wired in series with 150 Mohm SiCr load resistors. Bolometer 30 was shorted with no load resistor, and bolometer 31 was shorted with a load resistor in the circuit. In all cases, the signals were routed to JFET gates. Bolometers 1-16 were not wired. The array was clipped to the detector board with beryllium-copper clips and gold wedge bonded.

Other changes to the system include using only 3 of 4 G10 support tubes, only 2 of 4 manganin cables, avoidance of the surface mount connectors on the detector board, and using gold leaf between the INVAR base plate, detector board, heat strap, and GRT interfaces.

Sample IV curves for all bolometers at a single temperature:
IV for 5327 LH5 IV for 5327 LH5, origin

IV curves for a bolometer at multiple temperatures:
IV's for 5327 LH5 bolometer 24
In the above graph, the filled circles are measured data, and the lines are model fits with 4 free parameters.

The preliminary summary of bolometer parameters is as follows:

G0 test array 5327 LH5
Data at 328, 342, 389, 467, 651, 860, 962, 4127 mK
Updated 16 Apr 2000, 16:10 PDT

R = R0 exp(sqrt(T0/T))
G = G0 T^beta

            R0    T0     G0=G(1 K)          G(0.3 K)*  G(0.5 K)
bol./grp.  ohms    K    W/K^(beta+1)  beta     W/K       W/K
---------  ----  -----  ------------  ----  ---------  ---------
17  2D     1175  51.70    65.6 e-9    2.92
18  2D     1193  51.94    65.5 e-9    2.91   1.97 e-9   8.71 e-9
19  2D     1242  50.99    63.5 e-9    2.79
20  2C     1400  49.64     1.61e-9    1.24
21  2C     1271  51.22     1.70e-9    1.36   0.33 e-9   0.66 e-9
22  2C     1275  51.06     1.72e-9    1.35
23  2B     1354  50.20     1.65e-9    1.29
24  2B     1403  49.92     1.62e-9    1.28   0.35 e-9   0.67 e-9
25  2B     1304  50.94     1.67e-9    1.35
26  2A     1228  52.05    63.8 e-9    2.88
27  2A     1291  51.77    62.0 e-9    2.83   2.05 e-9   8.72 e-9
28  2A     1268  51.47    63.7 e-9    2.80
29  1A     1232  52.47    25.4 e-9    2.69   1.00 e-9   3.93 e-9
32  mem.   1243  52.25   128   e-9    2.40   7.12 e-9  24.3  e-9

* extrapolation; may not be accurate

Here is a Postscript summary of the bolometer recipes to associate with the group specification in the above table.