Version 2.0 -- 25 June 2000
C. Darren Dowell
Caltech, Mail Code 320-47
Pasadena, CA 91125
(626)395-6675 (office)
(626)395-2600 (lab)
(626)796-8806 (FAX)
cdd@submm.caltech.edu
Figure 1. Caltech interface substrate (alumina). The
3-element implant screening devices
were GE varnished to the substrate, and the 32-element bolometer arrays were
clipped to the substrate. Gold wedge bonds provided electrical connection
between the devices, load resistors, and substrate. Surface-mount connectors
were initially
used to interface with the cryostat wiring. However, they were replaced
with directly soldered wires for later measurements. The ceramic substrate
was enclosed in an INVAR box. A sheet of gold leaf and the force
of ~6 screws were used to thermally connect the INVAR and ceramic.
Figure 2. Device suspension and shielding. The ceramic/
INVAR package was suspended from the L4He coldplate with 3-4
fiberglass tubes
with A/L = 0.014-0.018 cm. The package was heat sunk to the fridge through an
annealed OFHC copper strap with A/L = 0.014 cm. A calibrated LakeShore GRT
was attached to the ceramic substrate with a screw and gold leaf, approximately
5 cm from the detectors. The package was shielded by a secondary
L4He
enclosure with blackened walls. With this configuration, the coldest observed
GRT measurement for the detector package was 0.325 K, during which the
L4He bath was at 1.5 K, the fridge was at 0.291 K, and the
extrapolated parasitic heat load on the detector package was 2-4 microW. The
JFET modules were located outside the inner shield.
Figure 3. Additional implant screening hardware. In order
to confirm the implant resistances, the 3-element devices in the
Goddard-selected
ceramic package were clipped to a copper block and installed directly
on the SHARC II fridge coldhead or in a separate L4He cryostat
(the purple 'Barney' Dewar).
All wires going into the Dewar pass through one or two RFI-filtered connectors, which contain 4 nF capacitors from each pin to the Dewar wall.
The wirebonds were 0.001 inch diameter 99.99% gold. With 64 each 3 mm long wire bonds (total A/L=0.001 cm), and assuming a 0.3 K thermal conductivity of 0.1 W/cm/K (Touloukian et al. 1970), the effective G of the wirebonds was 100 microW/K. We have ignored unknown boundary thermal resistance, which will reduce the effective G.
The ceramic substrate is approximately 7 cm x 7 cm x 0.06 cm, giving an A/L of 800 cm in the thickness (shortest) direction. Assuming a 0.3 K thermal conductivity of 11 microW/cm/K (Locatelli et al. 1976), the thermal conductance from the INVAR enclosure to the trace side of the ceramic board is G = 9000 microW/K. Again, we have ignored boundary thermal resistance.
The heat strap also limits the cooling of the detector wafer. During the cooldowns reported here, the fridge/heat strap combination was observed to have an effective G of 80 microW/K at the coldest temperatures.
Combining the effective G's of the wire bonds, ceramic board, and heat strap, we estimate an effective G of 40 microW/K for the link from the silicon die to the 3He fridge, three orders of magnitude greater than the G of the membrane devices. For a maximum 0.5 V bias across the 64 each 30 Mohm load resistors and detectors, the dissipated power can be at most 0.5 microW, leading to a temperature rise of at most 13 mK. For the 0.5 V bias, the detector package was observed to slowly warm a few mK. However, the measurements with the 0.5 V bias were performed quickly, using the thermal inertia of the detector package to keep the detector cool.
Table 1. Recipes for bolometers on G0 test
arrays.
Table 2. Recipes for bolometers on thermistor test arrays.
Table 3. Implant doses. Higher net density means greater
implant dose, which means lower cold resistance. Wafers 5347 and 5327 are
of interest for SHARCII and HAWC. (Wafer 5168 was not tested.)
Resistance measurements were made with a Keithley 616 Digital Electrometer borrowed from Goddard that allows current excitations from 1 pA to 1 microA stepped by factors of 10. The reported measurements were made with the highest current setting which did not cause significant self heating. The criterion was to allow no more than a 2% effect on the observed resistance.
A fit to the data with model R = R0 exp[(T0/T)1/2] yields:
Label ND R0 (ohms) T0 (K) ----- ---- --------- ------ 5250a 0.6 2570 81.8 5250b 0.6 3880 74.6 5327 0.7 1750 48.9 5347 0.75 2090 25.7Table 4. Fits for resistance of implant devices on Caltech ceramic board.
Figure 6. Measurements of thermistor samples in the
purple Dewar.
A fit to the data from the purple Dewar experiment only yields:
Label ND R0 (ohms) T0 (K) ----- ---- --------- ------ 5306 0.65 1670 67.3 5327 0.7 2100 42.7 5347 0.75 2380 23.2Table 5. Purple Dewar thermistor fits.
Label ND R0 (ohms) T0 (K) ----- ---- --------- ------ 5347 0.75 2430 22.5Table 6. SHARCII coldhead thermistor fit.
Figure 9. Measurements of 5347 thermistor samples. Two membrane
dies and three 32-element arrays were measured. (The thermistor array
measurements
were normalized to L/W = 1.45.) There is considerable variation from
sample to sample, which nearly covers the range from the HAWC target to the
SHARC II target. There is good agreement between the measurements for
membrane die A in the purple Dewar (green curve) and SHARC II (black curve).
Figure 10. Circuit used to measure IV curves. 16 JFETs
were available, so up to 16 bolometers were measured simultaneously. The
JFET gain was approximately 0.996. The bias voltage was slowly stepped
from 0 V to a nonzero value. The change in voltage from the JFET gives
the bolometer voltage, which yields the bolometer resistance since the
load resistance is known (to 5% accuracy).
Figure 11. Sample IV curves for all bolometers at a
single temperature.
Figure 12. IV curves for a bolometer at multiple
temperatures. The filled circles are measured data, and the lines are
model fits with 4 free parameters.
The bolometer parameters are derived from a minimization procedure in which the measured I's and V's are compared with a best-fit four-parameter model. The model is:
G0 test array 5327 LH5 Data at 328, 342, 389, 467, 651, 860, 962, 4127 mK Negative data included; 4127 mK duplicated Updated 1 May 2000, 16:30 PDT R = R0 exp(sqrt(T0/T)) G = G0 T^beta R* R R0 T0 0.3 K 0.5 K G0=G(1 K) G(0.3 K)* G(0.5 K) bol/grp ohms K Mohms Mohms W/K^(beta+1) beta W/K W/K ------- ---- ----- ----- ----- ------------ ---- --------- --------- 17 HG2D 1195 51.45 582 30.4 62.3 e-9 2.85 2.02 e-9 8.64 e-9 18 HG2D 1222 51.62 608 31.6 61.9 e-9 2.84 2.03 e-9 8.65 e-9 19 HG2D 1289 50.52 557 29.9 59.0 e-9 2.70 2.29 e-9 9.08 e-9 20 HG2C 1481 48.99 525 29.5 1.56e-9 1.18 0.38 e-9 0.69 e-9 21 HG2C 1321 50.76 589 31.4 1.66e-9 1.31 0.34 e-9 0.67 e-9 22 HG2C 1316 50.67 580 31.0 1.68e-9 1.31 0.35 e-9 0.68 e-9 23 HG2B 1412 49.70 549 30.2 1.61e-9 1.25 0.36 e-9 0.68 e-9 24 HG2B 1474 49.36 549 30.4 1.57e-9 1.23 0.36 e-9 0.67 e-9 25 HG2B 1363 50.43 582 31.3 1.62e-9 1.30 0.34 e-9 0.66 e-9 26 HG2A 1277 51.60 634 33.0 59.9 e-9 2.81 2.03 e-9 8.54 e-9 27 HG2A 1358 51.20 640 33.7 57.5 e-9 2.74 2.12 e-9 8.61 e-9 28 HG2A 1336 50.88 605 32.1 58.8 e-9 2.71 2.25 e-9 8.99 e-9 29 HG1A 1290 51.96 670 34.5 24.2 e-9 2.63 1.02 e-9 3.91 e-9 32 mem. 1299 51.76 658 34.1 202 e-9 2.31 12.5 e-9 40.7 e-9 * extrapolationTable 7. Four-parameter fits for G0 bolometer array 5327 LH5.
The fits in Table 7 are somewhat unsatisfactory, as evidenced by the systematic underestimate of the bolometer resistance at the lowest and highest temperatures (Figure 12). Better fits are obtained with a slightly different thermistor behavior, such as R = R0 exp[(T0/T)0.59]. Alternatively, we fit just the data at the colder temperatures to achieve better accuracy there.
Figure 13. IV curves for bolometer 24, with fit to low
temperature data only (T < 0.8 K).
G0 test array 5327 LH5 Data at 328, 342, 389, 467, 651 mK Updated 1 May 2000, 16:40 PDT R = R0 exp(sqrt(T0/T)) G = G0 T^beta R* R R0 T0 0.3 K 0.5 K G0=G(1 K) G(0.3 K)* G(0.5 K) bol/grp ohms K Mohms Mohms W/K^(beta+1) beta W/K W/K ------- ---- ----- ----- ----- ------------ ---- --------- --------- 17 HG2D 599 58.28 677 29.3 123 e-9 3.63 1.56 e-9 9.93 e-9 18 HG2D 539 59.68 720 29.9 135 e-9 3.72 1.53 e-9 10.2 e-9 19 HG2D 590 58.26 665 28.8 129 e-9 3.61 1.67 e-9 10.6 e-9 20 HG2C 581 58.39 665 28.7 2.44e-9 1.72 0.31 e-9 0.74 e-9 21 HG2C 580 59.08 721 30.5 2.46e-9 1.78 0.29 e-9 0.72 e-9 22 HG2C 589 58.76 705 30.1 2.47e-9 1.76 0.30 e-9 0.73 e-9 23 HG2B 587 58.51 682 29.3 2.45e-9 1.76 0.29 e-9 0.72 e-9 24 HG2B 597 58.46 689 29.6 2.41e-9 1.75 0.29 e-9 0.72 e-9 25 HG2B 596 58.78 715 30.5 2.40e-9 1.77 0.28 e-9 0.70 e-9 26 HG2A 632 58.62 744 31.9 118 e-9 3.60 1.55 e-9 9.73 e-9 27 HG2A 636 58.80 765 32.6 120 e-9 3.60 1.57 e-9 9.90 e-9 28 HG2A 627 58.44 722 31.1 124 e-9 3.58 1.67 e-9 10.4 e-9 29 HG1A 630 59.22 796 33.6 42.6 e-9 3.30 0.80 e-9 4.33 e-9 32 mem. 650 58.62 765 32.8 610 e-9 3.54 8.60 e-9 52.4 e-9 * extrapolationTable 8. Four-parameter fits for array 5327 LH5, using only data at T < 0.8 K.
The fits using only the low temperature data give 0.3 K resistances which are ~20% higher, 0.5 K resistances which are ~2% lower, 0.3 K G's which are ~25% lower, and 0.5 K G's which are ~15% higher.
The derived values of beta are somewhat surprising. The devices with only silicon legs have measured beta's of 1.3 - 1.8, while beta's of 3 (the value expected for a dielectric) were measured in the 12 micron silicon bolometers (e.g., Wang et al. 1996). The speculation is that the thermal conductance is dominated by the heavily doped traces, which have a metallic beta of 1 (H. Moseley, private communication).
The devices with metal on the legs, on the other hand, have steep beta's of 2.7 - 3.6. While normal metals should give a beta of 1, the aluminum metallization is superconducting at these temperatures and may have a much more steep beta.
The addition of 100 micron wide metal to the strap has essentially no effect on the thermal conductance. A bolometer with the design HGUNIT.2D (metal) has the same G as HGUNIT.2A (no metal), and HGUNIT.2C (metal) has the same G as HGUNIT.2B (no metal).
Figure 14. Sample IV curves for all bolometers at a single
temperature.
Figure 15. Sample IV curves for a single bolometer at
multiple temperatures.
G0 test array 5347 LH1 Data at 330, 440, 597, 742, 985, 4243 mK (pixels 1-16) Data at 329, 448, 604, 744, 974, 4261 mK (pixels 17-32) Updated 1 May 2000, 16:50 PDT R = R0 exp(sqrt(T0/T)) G = G0 T^beta R* R R0 T0 0.3 K 0.5 K G0=G(1 K) G(0.3 K)* G(0.5 K) bol/grp ohms K Mohms Mohms W/K^(beta+1) beta W/K W/K ------- ---- ----- ----- ----- ------------ ---- --------- --------- 01 mem. 1844 26.25 21.3 2.59 23.3 e-9 0.55 12.1 e-9 15.9 e-9 02 TH.1 1290 25.58 13.2 1.65 1.07e-9 1.32 0.22 e-9 0.43 e-9 03 TH.1 1365 25.51 13.8 1.73 1.20e-9 1.19 0.29 e-9 0.53 e-9 04 TH.1 1401 25.31 13.7 1.72 0.94e-9 1.15 0.24 e-9 0.42 e-9 05 HG3D 1953 26.94 25.5 3.01 24.4 e-9 1.97 2.28 e-9 6.23 e-9 06 HG3D 1956 27.16 26.5 3.11 24.1 e-9 1.96 2.28 e-9 6.19 e-9 07 HG3D 2147 27.11 28.9 3.39 19.4 e-9 1.77 2.30 e-9 5.69 e-9 08 HG3C 2127 26.93 27.7 3.27 2.24e-9 1.20 0.53 e-9 0.98 e-9 09 HG3C 1844 27.58 26.9 3.10 2.66e-9 1.38 0.51 e-9 1.02 e-9 10 HG3C 1898 27.05 25.2 2.97 2.61e-9 1.35 0.51 e-9 1.02 e-9 11 HG3B 1934 28.40 32.5 3.63 24.2 e-9 2.03 2.10 e-9 5.93 e-9 12 HG3B 1908 28.51 32.7 3.63 25.1 e-9 2.06 2.10 e-9 6.02 e-9 13 HG3B 1869 28.76 33.4 3.68 27.8 e-9 2.14 2.11 e-9 6.31 e-9 14 HG3A 1840 28.67 32.4 3.58 4.08e-9 1.44 0.72 e-9 1.50 e-9 15 HG3A 1951 28.41 32.9 3.66 3.80e-9 1.42 0.69 e-9 1.42 e-9 16 HG3A 1924 28.37 32.2 3.59 3.89e-9 1.41 0.71 e-9 1.46 e-9 17 HG2D 1893 28.73 33.7 3.71 38.4 e-9 2.63 1.62 e-9 6.20 e-9 18 HG2D 1899 28.90 34.8 3.80 38.7 e-9 2.66 1.57 e-9 6.12 e-9 19 HG2D 1981 28.82 35.8 3.93 36.7 e-9 2.65 1.51 e-9 5.85 e-9 20 HG2C 2025 28.51 34.7 3.85 1.17e-9 1.35 0.23 e-9 0.46 e-9 21 HG2C 1839 28.97 34.1 3.72 1.31e-9 1.43 0.23 e-9 0.49 e-9 22 HG2C 1920 28.91 35.2 3.85 1.26e-9 1.40 0.23 e-9 0.48 e-9 23 HG2B 1957 29.47 39.4 4.22 1.21e-9 1.37 0.23 e-9 0.47 e-9 24 HG2B 2038 29.15 38.9 4.22 1.17e-9 1.34 0.23 e-9 0.46 e-9 25 HG2B 1929 29.84 41.4 4.37 1.25e-9 1.40 0.23 e-9 0.47 e-9 26 HG2A 1871 30.40 44.0 4.55 42.4 e-9 2.76 1.53 e-9 6.26 e-9 27 HG2A 2019 29.93 44.0 4.63 37.3 e-9 2.62 1.59 e-9 6.07 e-9 28 HG2A 1952 30.41 46.0 4.76 40.0 e-9 2.69 1.57 e-9 6.20 e-9 29 HG1A 1836 31.16 49.0 4.92 20.0 e-9 2.74 0.74 e-9 2.99 e-9 30 HG1A 1788 31.95 54.2 5.30 20.7 e-9 2.80 0.71 e-9 2.97 e-9 31 HG1A 1927 31.34 52.9 5.29 18.1 e-9 2.68 0.72 e-9 2.82 e-9 32 mem. 1814 32.12 56.5 5.49 72.9 e-9 1.58 10.9 e-9 24.4 e-9 * extrapolationTable 9. Four-parameter fits for G0 bolometer array 5347 LH1.
Although the measured values of G make qualitative sense, the scaling with leg geometry is not what was expected. This will be discussed further in the Conclusions.
Electrometer Table 9 Fit Resistance Resistance Bolometer Mohm Mohm --------- ------------ ----------- 1 15.6 14.4 9 19 18.0 12 23 21.7 15 22 21.8 24 28 25.7 32 36.4 36.6Table 10. Resistance gradient in G0 array 5347 LH1.
A radiation gradient as the cause of the resistance can be ruled out since the membrane (high G) devices and bolometers (low G) are affected about the same. WHAT ABOUT A TEMPERATURE GRADIENT???
Figure 17. Noise spectra for bolometers 22-21 on 5347 G0
array LH1. The bolometers are of type HGUNIT2C. The observed noise has been
divided by the amplifier gain (except the
low pass filtering) and by 21/2 to show the noise for a single
bolometer referred to the JFET input. The 1/f noise in the 40 mV and 157 mV
spectra is most likely due to operation of the JFETs at a non-optimal
temperature. The short horizontal lines at right
show the noise predicted from the IV curve fits in the ~30 Hz region. The
predictions are in good agreement with the observations. The calculated
electrical NEP for the optimal 40 mV bias is 6.1x10-17
W s1/2.
For the time constant measurements, the bolometers were driven using a bias waveform with a square profile switching between two positive levels 8 mV apart. The output of the JFET was sent to an SR560 preamplifier with a gain of 100 and then to an oscilloscope. DC coupling was used on both instruments; a DC level from a power supply was subtracted to bring the signal on scale. Low-pass filtering was applied with the SR560, but the cutoff frequency (typically 3 kHz) was chosen so that only high frequency noise was eliminated and the shape of the waveform was preserved.
Additional filtering of the bolometer waveform was caused by the bias waveform filtering (cutoff frequency at 6800 Hz) and the RFI connectors (cutoff frequency at 20 kHz, assuming a 1000 ohm JFET output impedance). These cutoff frequencies are high enough to be irrelevant.
Figure 19. Typical oscilloscope trace during time constant
measurement. Shown is the output of the JFET for bolometer 24 of the 5347
G0 test array, multiplied by a gain of 100. The bias frequency was
8 Hz, and the settling time was 3.6 msec.
The overshoot is caused by the high-frequency impedance of the bolometer being larger than the low-frequency impedance (Mather 1982). Stated another way, the dynamic circuit model of a bolometer contains an effective inductance, which creates a voltage spike at the transition of the bias. Our interpretation is that the settling time of the voltage following the bias transition is the true detector time constant (taue in the notation of Mather 1982).
For the measured quantities, we looked at the transition caused by the positive change in the bias. We recorded:
calc. observ. bias calc G calc T V(o.s.) t(o.s.) V(step) V(step) t(settle) bol. grp. mV nw/K K mV msec mV mV msec ---- ---- ---- ------ ------ ------- ------- ------- ------- --------- 9 3C 25 0.59 0.34 0.12 2.2 2.9 2.7 1.5 9 3C 49* 0.64 0.36 0.42 1.1 1.4 1.4 1.6 9 3C 98 0.76 0.40 0.70 0.52 0.34 0.40 1.2 9 3C 196 0.98 0.48 0.61 0.28 0.02 0.06 0.71 15 3A 39 0.84 0.34 0.19 1.8 2.2 2.2 1.6 15 3A 78?* 0.95 0.38 2.8? 0.92 0.90 6.6? 1.1 15 3A 157 1.20 0.44 0.69 0.31 0.12 0.18 0.61 15 3A 314 1.60 0.54 0.47 0.24 -0.02 0.05 0.31 18 2D 49 2.1 0.34 N.A. N.A. 2.7 2.6 0.38 18 2D 98* 2.5 0.36 0.20 0.8 1.5 1.5 0.68 18 2D 196 3.6 0.41 0.46 0.31 0.45 0.53 0.36 18 2D 392 5.6 0.48 0.34 0.22 0.12 0.22 0.16 24 2B 20 0.28 0.34 0.39 2.4 2.6 2.7 4.5 24 2B 39* 0.31 0.37 1.1 1.3 1.2 1.1 3.6 24 2B 78 0.38 0.43 1.2 0.6 0.17 0.11 2.3 24 2B 157 0.50 0.53 0.74 0.4 -0.03 -0.13 1.3 * optimum bias for NEP ? apparent errors in measurement, probably bias amplitude settingTable 11. Time constant measurements for G0 array 5347 LH1 operated at 0.328 K.
Thermistor test array 5347 RH7 Data at 330, 451, 610, 755, 971, 4190 mK Updated 1 May 2000, 16:55 PDT; corrected grp labels 25 June 2000 R = R0 exp(sqrt(T0/T)) G = G0 T^beta R* R R0 T0 0.3 K 0.5 K G0=G(1 K) G(0.3 K)* G(0.5 K) bol/grp ohms K Mohms Mohms W/K^(beta+1) beta W/K W/K ------- ---- ----- ----- ----- ------------ ---- --------- --------- 09 TH.4 1622 35.93 91.8 7.79 1.53e-9 1.67 0.20 e-9 0.48 e-9 10 TH.4 1658 35.50 87.9 7.57 1.52e-9 1.64 0.21 e-9 0.49 e-9 11 TH.4 1622 35.65 88.0 7.54 1.54e-9 1.69 0.20 e-9 0.48 e-9 12 TH.3 1800 34.29 79.1 7.11 1.42e-9 1.49 0.24 e-9 0.51 e-9 13 TH.3 1733 34.60 80.0 7.10 1.48e-9 1.53 0.23 e-9 0.51 e-9 14 TH.3 1726 34.57 79.3 7.05 1.49e-9 1.52 0.24 e-9 0.52 e-9 15 TH.3 1800 34.52 82.0 7.31 1.44e-9 1.51 0.23 e-9 0.51 e-9 16 TH.3 1802 34.41 80.7 7.22 1.44e-9 1.50 0.24 e-9 0.51 e-9 17 TH.2 1638 36.36 99.0 8.28 1.50e-9 1.57 0.23 e-9 0.51 e-9 18 TH.2 1668 36.19 98.2 8.26 1.49e-9 1.56 0.23 e-9 0.51 e-9 19 TH.2 1727 36.17 101 8.53 1.43e-9 1.56 0.22 e-9 0.48 e-9 20 TH.2 1693 36.09 98.2 8.29 1.46e-9 1.56 0.22 e-9 0.50 e-9 21 TH.2 1578 36.79 102 8.38 1.55e-9 1.62 0.22 e-9 0.50 e-9 22 TH.1 1096 33.11 40.0 3.75 1.42e-9 1.45 0.25 e-9 0.52 e-9 23 TH.1 1106 33.47 42.8 3.95 1.37e-9 1.42 0.25 e-9 0.51 e-9 24 TH.1 1108 33.03 40.0 3.75 1.37e-9 1.42 0.25 e-9 0.51 e-9 * extrapolationTable 12. Four-parameter fits for thermistor test array 5347 RH7.
There is no strong resistance gradient across the RH7 thermistor device, as confirmed by the following 0.330 K measurements with an electrometer:
Electrometer Resistance Bolometer Mohm --------- ------------ 1 42.8 6 44 31 44 32 35.0Table 13. Electrometer resistance measurements for thermistor test array 5347 RH7.
Figure 21. Noise spectra for bolometers 10-9 on 5347 thermistor
test array RH7. The bolometers are of type THUNIT2B.4 and have small
thermistors. The veritical display range has been enlarged by a factor of
2 compared to Figure 20.
Figure 23. Low-frequency noise spectra for bolometers 23-22 and 10-9
on 5347 thermistor test array RH7. The noise has been divided by
21/2 and corrected for the amplifier gain to express the noise
for a single bolometer. The predictions are based on the IV
curves and the 1/f noise model of Han et al. (1998). For the bolometers with
full-area thermistors, an NEP of 2x10-16 W s1/2
was observed at 0.01 Hz, and 9x10-17 W s1/2 at 0.1 Hz.
calc. observ. bias calc G calc T V(o.s.) t(o.s.) V(step) V(step) t(settle) bol. grp. mV nw/K K mV msec mV mV msec ---- ---- ---- ------ ------ ------- ------- ------- ------- --------- 10 TH.4 50* 0.34 0.40 1.3 1.3 0.82 0.80 3.6 10 TH.4 100 0.47 0.49 1.1 0.36 -0.044 -0.025 2.1 23 TH.1 50* 0.36 0.39 1.3 1.2 0.73 0.70 4.2 23 TH.1 100 0.45 0.46 1.1 0.58 0.035 0.019 2.8 * optimum bias for NEPTable 14. Time constant measurements for thermistor test array 5347 RH7 operated at 0.331 K.
At these temperatures, the size of the thermistor only plays a small role in the time constant and heat capacity -- 30% at most.
before after before after before after before after R* R* R R 0.3 K 0.3 K 0.5 K 0.5 K G(0.3 K)* G(0.3 K)* G(0.5 K) G(0.5 K) bol/grp Mohms Mohms Mohms Mohms W/K W/K W/K W/K ------- ----- ----- ----- ----- --------- --------- --------- --------- 19 TH.2 101 115 8.53 8.59 0.22 e-9 0.18 e-9 0.48 e-9 0.49e-9 21 TH.2 102 117 8.38 8.47 0.22 e-9 0.18 e-9 0.50 e-9 0.51e-9 23 TH.1 42.8 46.9 3.95 4.07 0.25 e-9 0.21 e-9 0.51 e-9 0.48e-9 24 TH.1 40.0 43.5 3.75 3.94 0.25 e-9 0.21 e-9 0.51 e-9 0.49e-9 29 2B NA 80.5 NA 6.44 NA 0.20 e-9 NA 0.55e-9 30 2B NA 80.0 NA 6.55 NA 0.20 e-9 NA 0.53e-9 32 mem. NA 68.5 NA 6.00 NA 9.12 e-9 NA 33.2 e-9 * extrapolationTable 15. Comparison of resistance and thermal conductance measurements before and after application of Pd/Au absorbing film onto thermistor test array RH7.
before after bias calc G calc T t(settle) t(settle) bol. grp. mV nw/K K msec msec ---- ---- ---- ------ ------ --------- --------- 21 TH.2 76 0.44 0.46 15 23 TH.1 24 0.32 0.36 16 23 TH.1 50 0.36 0.39 4.2 19 23 TH.1 76 0.42 0.43 16 23 TH.1 100 0.45 0.46 2.8 16 30 2B 76 0.44 0.45 15Table 16. Comparison of time constants before and after application of absorbing film. The 'before' measurements had an operating temperature of 0.331 K, and the 'after' measurements had an operating temperature of 0.345 K.
Figure 25. Oscilloscope trace during time constant measurements
of bolometer 23 on thermistor array 5347 with an absorbing film. Compare
with Figure 24, and note the change in horizontal scale (msec/division).
The most perplexing aspect of the detector measurements is the behavior of the thermal conductances, summarized in the table below:
tors. bar tors. bar G(0.3 K)* G(0.5 K) therm. struct. wafer/sample silicon W/L metal W/L nW/K nW/K ---------------- ------------ ----------- --------- ---------- ----------- HG2B, HG2C, TH2B 5327 G0 LH5 0.038 0 0.28- 0.38 0.66- 0.74 HG2B, HG2C, TH2B 5347 G0 LH1 0.038 0 0.22- 0.29 0.42- 0.53 HG2B, HG2C, TH2B 5347 th RH7 0.038 0 0.20- 0.25 0.48- 0.52 HG3C 5347 G0 LH1 0.068 0 0.51- 0.53 0.98- 1.02 HG3A 5347 G0 LH1 0.068 0.136^ 0.69- 0.72 1.41- 1.50 HG1A 5327 G0 LH5 0.022 0.022 0.80- 1.02 3.91- 4.33 HG1A 5347 G0 LH1 0.022 0.022 0.71- 0.74 2.82- 2.99 HG2A, HG2D 5327 G0 LH5 0.038 0.038 1.53- 2.29 8.54-10.6 HG2A, HG2D 5347 G0 LH1 0.038 0.038 1.51- 1.62 5.85- 6.26 HG3B, HG3D 5347 G0 LH1 0.068 0.068 2.10- 2.30 5.69- 6.31 membrane 5327 G0 LH5 large large 8.60-12.5 40.7 -52.4 membrane 5347 G0 LH1 large large 10.9 24.4 * extrapolation ^ only on cold half of torsion barTable 17. Summary of thermal properties of HGUNIT's and THUNIT's.
As noted in the section "Results for 5327 G0 Test Array LH5", the G's for metal-free legs go as ~T1.5, and the G's for the metallized legs go as ~T3. We have offered an explanation for the unexpected temperature dependence in that section.
The HGUNIT.2B/C, HGUNIT.3C, and HGUNIT.3A devices on 5347 G0 array LH1 can be used to estimate the dependence of G on torsion bar width and length for devices with no metal; we calculate a W1.06L-0.45 dependence at 0.3 K and W1.03L-0.53 at 0.5 K. (We have assumed that the metal half-way up the torsion bar on HGUNIT.3A is sufficient to cool that part of the bar to the base operating temperature.) In macroscopic/high temperature situations, we expect W1 and L-1, so the observed L dependence is surprising.
For the metallized devices (HGUNIT.1A, HGUNIT.2A/D, and HGUNIT.3B/D), the situation is even more surprising. The range of torsion bar lengths is not large, so we fit for the dependence of G on (W/L) together and calculate (W/L)0.99 at 0.3 K and (W/L)0.64 at 0.5 K, dominated presumably by width dependence. This is quite far from the expected (W/L)2 dependence. It is possible that the 0.5 K measurements of the metallized bolometers especially are contaminated by heating effects in the silicon frame and/or detector package.
The following table summarizes the expected operating conditions of the SHARCII bolometers as designed above:
parameter value units ----------------------- ---------- --------- operating temperature 0.32 K T0 40 K R0 1300 ohm thermistor volume 2.7e-7 cm^3 G0 1.9 W K^-2.35 beta (G = G0 T^beta) 1.35 R (load resistor) 200 Mohm T (load resistor) 2.0 K background power 75 pW NEP (background) 4.4e-16 W s^1/2 voltage noise (amp.) 5 e-9 V s^1/2 V (bias) 500 mV V (bolometer) 19.4 mV I 2.4 nA R (bolometer) 8.1 Mohm Z (bolometer) 2.8 Mohm T (bolometer) 0.52 K G 0.79 nW/K S 1.3e+8 V/W NEP (detector, 0.01 Hz) 2.8e-16 W s^1/2 NEP (detector, 0.1 Hz) 1.4e-16 W s^1/2 NEP (detector, 1 Hz) 1.2e-16 W s^1/2 NEP (detector, 10 Hz) 1.2e-16 W s^1/2 det. volt. noise, 10 Hz 16 e-9 V s^1/2 time constant (elec.) 1 msecTable 18. SHARCII bolometer characteristics, updated from Moseley et al. (1999).
Locatelli, M., Arnaud, D., & Routin, M. 1976, *journal unknown*, "Thermal conductivity of some insulating materials below 1K"
Mather, J. C. 1982, Applied Optics 21, 1125, "Bolometer noise: nonequilibrium theory"
Moseley, H. 1999, memo dated July 29, "Detector Design for HAWC"
Moseley, S. H, Dowell, C. D., Allen, C, & Phillips, T. G. 1999, in Imaging at Radio Through Submillimeter Wavelengths, ed. J. Mangum & S. Radford, ASP Conf. Ser., "Semiconducting Pop-Up Bolometers for Far-Infrared and Submillimeter Astronomy"
Toloukian, Y. S., Powell, R. W., Ho, C. Y., & Klemens, P. G. 1970, Thermal Properties of Matter, Volume 1, Thermal Conductivity -- Metallic Elements and Alloys, IFI/Plenum: New York
Wang, N., et al. 1996, Applied Optics, 35, 6629, "Characterization of a submillimeter high-angular-resolution camera with a monolithic silicon bolometer array for the Caltech Submillimeter Observatory"