Testing the QMMIC on a Probe Station

Probe Station


Warnings

WARNING: It is extremely easy to destroy the QMMIC while probing. Bumping any of the capacitors or air bridges will destroy the chip.

WARNING: It is vitally important that the QMMIC be kept in a safe place until the probe station is completely set up and ready to go.

WARNING: Practice probing a dead chip before taking a good chip out of the drawer.

WARNING: Don't try to pick up the chip with bent or dirty tweezers. The chip will slip out of the tweezers and get scratched or lost or otherwise broken.

WARNING: Don't bump the probes, probe station, or table top while the chip is being probed.

WARNING: Be sure to examine all probe tips and test all the positioners before taking out the chip. In particular, make sure that the positioners move smoothly without catching or jerking. If they do not move smoothly, fix them before taking the chip out of the drawer!


Detailed Instructions

  1. Make sure the QMMIC chip is safely put away in a Lista drawer. Don't get it out until the probe station is completely ready!
  2. Inspect the probe positioners carefully. Move all screws to approximately their center positions. Make sure the motion is smooth.
  3. Install the 150 micron microwave probes. Inspect the DC probe needles, and replace any that have excessively bent or dull tips. Then install the DC probes.
  4. Position the DC probes tips to be reasonably close to where they will need to be. Make sure that they are close enough that the last little bit can be adjusted entirely with the micro positioners. Make sure that the positioners will have enough downward travel left to contact the chip.
  5. Probing the QMMIC

  6. Hook up the bias supplies. The grounds can be connected through the network analyzer. You will need
  7. Set all bias voltages to 0. Double check the connections. Make sure none of the probes is shorted to anything. Check each supply to make sure no current flows for non-zero voltages.
  8. Calibrate the network analyzer. The test power should be set to about -40 dBm. Test the cal to make sure it looks good. Get a blank floppy disk ready to store data.
  9. Find a good pair of tweezers. Make sure the tips aren't bent. Clean the tips carefully with acetone.
  10. Place a dead MMIC chip on a piece of circuit board. Put the circuit board on the probe station. Position the microwave probes just above the CPW pads, but do not contact them yet. Very carefully position the DC probes just above the DC bias probes, but don't contact them yet.
  11. Probing the QMMIC

  12. Only after all probe tips are correctly positioned just above the pads, contact the microwave probe tips, then the DC probe tips.
  13. Now that you have gotten some practice with the dead chip, and gotten all the probes perfectly lined up, carefully raise the DC probes about a millimeter, one at a time, then raise the RF probes. Put away the dead chip and replace it with a good chip. Lower and contact the RF probe tips as before, then contact the DC probe tips.
  14. Turn the drain voltages up to 0.5 V. There should be a little bit of current on the drains, very roughly about 1 mA. Turn the voltages up to about 1.0 V. Then check each gate, one at a time. Turning a gate voltage up by minus a few tenths of a volt should decrease the corresponding drain current. If it doesn't, turn the gate voltage back to zero and check all wiring, inspect the probe positioning under the microscope, etc. until the problem has been fixed.
  15. By now, S21 should have some gain on the network analyzer. Adjust the drain and gate voltages as desired. Try to not let the drain current get larger than 6 or 7 mA, as there is no reason to do so.
  16. As soon as testing is finished and all data has been stored, turn all the bias supplies to 0. Carefully raise each DC probe one at a time, being extremely careful to not scratch the chip. Then raise the microwave probes. Put the chip away before doing anything else.


Understanding the Measured Results

The results may have undesirable features such as gain and reflection peaks. These may be related to substrate modes and/or grounding problems. These features may change or go away when the chip is mounted in a housing, and can usually be reduced by adequately grounding both sides of the mounted chip with several wire bonds.

It should also be noted that the transconductances (and thus the gain) will increase dramatically as the chip is cooled.